DO Starter Benchtop Meter
HI2640 and HI2641 enable fast, accurate measurements utilizing the Hanna Instruments® HI764080 digital dissolved oxygen probe with an integral temperature sensor. The probe has a unique serial number and, once connected, is automatically identified by the meter.
Both meters measure dissolved oxygen in ppm and % saturation while the HI2641 adds data logging capabilities. The intuitive design simplifies configuration, calibration, measurement, data logging and transfer (to a USB thumb drive or computer). HI2640 and HI2641 benchtop meters offer a Basic Mode that streamlines measurement configuration and is useful for routine applications.
Additionally, HI2641 can be used in Standard Operating Mode, where all features and capabilities are enabled.
- Automatic parameter recognition
- Choice of measurement unit:
- ppm (mg/L)
- % saturation
- Simplified operation for routine measurements


HI2640 Features
Data Logging: No
USB-C PC Transfer: No


HI2641 Features
Data Logging: Yes
USB-C PC Transfer: Yes
Designed for simple and intuitive operation
The color LCD display with a capacitive touch keypad and clear on-screen text guides users through setup, calibration, and measurement with ease. Standard and Basic operating modes allow the interface and available functions to be tailored to the user’s experience level or application needs, making the instrument ideal for both experienced professionals and those who prefer a simple, streamlined workflow.
- Intuitive, guided user interfaces
- Auto sensor recognition
- Reduced errors and shorter learning time




Accurate, reliable results with HANNA Digital Smart Probe
The meters measure dissolved oxygen through its unique digital electrode. The HI2640 and HI2641 include the HI764080 digital polarographic dissolved oxygen probe with integral temperature sensor. This digital electrode is auto-recognized, providing sensor type, calibration data, and a serial number when connected by an easy to plug-in 3.5 mm connector.
- Direct signal processing for noise-free measurements
- Automatic sensor recognition
- Store calibration specific data from the last calibration
- Built with materials suitable for use in chemical analysis
- Integral temperature sensor
- 3 mm jack termination
- Connectors are sealed against moisture and corrosion and are more reliable than traditional BNC connectors.
- Unique serial ID for probe traceability
Watch & Discover the HI2600 Series
A quick look at its design and ease of use.
Specifications
| HI2640 and HI2641 | Using the Dissolved Oxygen probe HI764080 |
| DO Range |
0.00 to 45.00 ppm (mg/L) 0.0 to 300.0% |
| DO Resolution |
0.01 ppm (mg/L) 0.1% |
| DO Accuracy at 25 °C / 77 °F | ±1.5% of reading ±1 digit |
| Temperature Range |
-20.0 to 120.0 °C * (-4.0 to 248.0 °F) |
| Temperature Resolution |
0.1 °C (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F |
±0.5 °C (0.9 °F) |
| DO Calibration |
One or two points at: 0% (HI7040) 100% (water-saturated air) |
| DO Pressure Compensation |
Automatic 450.0 to 850.0 mmHg |
| DO Pressure Compensation Resolution | 0.1 mmHg |
| DO Salinity Compensation | 0 to 40 g/L |
| DO Salinity Compensation Resolution | 1 g/L |
| Temperature Compensation | 0.0 to 50.0 °C (32.0 to 122 °F) |
| Logging (HI2641 only) |
Up to 1000 records organized in: • Manual log-on-demand (max 200 logs) • Manual log-on-stability (max 200 logs) • Interval logging (max 600 logs / 100 lots)** |
| Notes |
* Temperature limits will be reduced to actual probe limits. ** When operating in Standard Mode only! |
| PC Interface (HI2641 only) | USB-C |
| Power Supply | USB Type C (5 VDC) |
| Environment |
0 to 50 °C (32 to 122 °F) Maximum 95% RH non-condensing |
| Dimensions | 205 x 160 x 77 mm (8.0 x 6.2 x 3.0 inches) |
| Weight | Approximately 0.85 kg (1.87 lbs.) |