HI2630 and HI2631 enables fast, accurate measurements utilizing the Hanna Instruments® HI763100 digital conductivity electrode with an integral temperature sensor. The probe has a unique serial number and, once connected, is automatically identified by the meter. Both meters measure conductivity (EC) and TDS while HI2631 adds salinity measurement and calibration along with data logging capabilities. The intuitive design simplifies configuration, calibration, measurement, data logging and transfer (to a USB thumb drive or computer). HI2630 and HI2631 benchtop meters offer a Basic Mode that streamlines measurement configuration and is useful for routine applications. Additionally, HI2631 can be used in Standard Operating Mode, where all features and capabilities are enabled

  • Automatic parameter recognition
  • Choice of measurement unit:
    • Conductivity: μS/cm, mS/cm
    • TDS: ppm (mg/L), g/L
  • Simplified operation for routine measurements

HI2630 Features

Salinity Measurement: No 
Data Logging: No  
USB Data Transfer: No 

HI2631 Features

Salinity Measurement: Yes
Data Logging: Yes
USB Data Transfer: Yes

Designed for simple and intuitive operation

 

The color LCD display with a capacitive touch keypad and clear on-screen text guides users through setup, calibration, and measurement with ease. Standard and Basic operating modes allow the interface and available functions to be tailored to the user’s experience level or application needs, making the instrument ideal for both experienced professionals and those who prefer a simple, streamlined workflow.

  • Intuitive, guided user interfaces
  • Auto sensor recognition
  • Reduced errors and shorter learning time

Accurate, reliable results with HANNA Digital Smart Probe

 

These meters measure EC, TDS and Salinity (HI2631 only) through its unique digital electrode. HI763100 is a digital four-ring conductivity probe with integral temperature sensor. This probe is auto-recognized, providing sensor type, calibration data, and a serial number when connected by an easy to plug-in 3.5 mm connector.

  • Direct signal processing for noise-free measurements
  • Automatic sensor recognition
  • Store calibration specific data from the last calibration
  • Built with materials suitable for use in chemical analysis
  • Integral temperature sensor
  • 3 mm jack termination
    • Connectors are sealed against moisture and corrosion and are more reliable than traditional BNC connectors.
  • Unique serial ID for probe traceability

Watch & Discover the HI2600 Series

A quick look at its design and ease of use.

Specifications

HI2630 and HI2631 Using the four-ring conductivity probe
EC Range 0.00 to 29.99 μS/cm
30.0 to 299.9 μS/cm
300 to 2999 μS/cm
3.00 to 29.99 mS/cm
30.0 to 200.0 mS/cm
Up to 500.0 mS/cm absolute conductivity ‡
EC Resolution 0.01 μS/cm
0.1 μS/cm
1 μS/cm
0.01 mS/cm
0.1 mS/cm
EC Accuracy at 25 °C / 77 °F ±1% of reading
(±0.05 μS/cm or 1 digit, whichever is greater)
TDS Range 0.00 to 14.99 ppm (mg/L)
15.0 to 149.9 ppm (mg/L)
150 to 1499 ppm (mg/L)
1.50 to 14.99 g/L
15.0 to 100.0 g/L
Up to 400.0 g/L (with 0.80 factor) absolute TDS***
TDS Resolution 0.01 ppm
0.1 ppm
1 ppm
0.01 g/L
TDS Accuracy at 25 °C / 77 °F ±1% of reading
(±0.03 ppm or 1 digit, whichever is greater)
Salinity* Range 0.0 to 400.0 % NaCl
2.00 to 42.00 PSU
0.00 to 80.00 g/L
Salinity* Resolution 0.1% NaCl
0.01 PSU
0.01 g/L
Salinity* Accuracy at 25 °C / 77 °F ±1% of reading
Temperature Range −20.0 to 120.0 °C**
(−4.0 to 248.0 °F)
Temperature Resolution 0.1 °C (0.1 °F)
Temperature Accuracy at 25 °C / 77 °F ±0.5 °C (0.9 °F)
EC Calibration Cell factor calibration
Six standards available:
84 µS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm
One point offset: 0.00 μS/cm
Salinity Calibration* (% NaCl only) Single point with HI7037 standard
Temperature Compensation • Automatic, −5 to 100 °C (23 to 212 °F)
• NoTC, can be selected to measure absolute conductivity
Conductivity Temperature Coefficient* 0.00 to 6.00 %/°C (EC and TDS only)
(1.90 %/°C default)
TDS Factor* 0.40 to 0.80 (0.50 default)
DO Salinity Compensation 0 to 40 g/L
Logging (HI2631 only) Up to 1000* (400) records organized in:
• Manual log-on-demand (max 200 logs)
• Manual log-on-stability (max 200 logs)
• Interval logging (max 600 logs / 100 lots)*
Notes *When operating in Standard Mode in HI2631 only!
** Temperature limits will be reduced to actual probe limits.
*** Absolute conductivity (TDS) is the conductivity value without temperature compensation.
PC Interface (HI2631 only) USB-C
Power Supply USB Type C (5 VDC; 500 mA)
Environment 0 to 50 °C (32 to 122 °F)
Maximum 95% RH non-condensing
Dimensions 205 x 160 x 77 mm (8.0 x 6.2 x 3.0 inches)